docs: update Combo8 firmware fixes
This commit is contained in:
@@ -2,6 +2,11 @@
|
||||
title: FAQ
|
||||
keywords: LogicAnalyzer, debugger, link, RISCV, tool
|
||||
update:
|
||||
- date: 2026-08-12
|
||||
version: v0.2
|
||||
author: Sipeed
|
||||
content:
|
||||
- Updated CH7 floating-waveform and DAPLink serial troubleshooting
|
||||
- date: 2023-09-12
|
||||
version: v0.1
|
||||
author: lxo
|
||||
@@ -30,9 +35,9 @@ The following lists some common problems according to the form of Q&A, please tr
|
||||
|
||||
**A:**This may be the result of unstable contact that causes the device to disconnect, try to re-plug the device and then reconnect it to solve the problem.
|
||||
|
||||
**Q:**When using 8-channel sampling, it was found that waveforms also appeared when the D7 channel was not connected.
|
||||
**Q:**Why is the floating CH7 waveform noisy during 8-channel acquisition?
|
||||
|
||||
**A:**This is a problem to be solved, but this problem will not affect the sampled waveforms, and the waveform can still be acquired normally after connecting an external signal to the D7 channel.
|
||||
**A:**This issue is fixed in `slogic_combo8_pack_202608121500.bin`. [Update to this release or a newer one](./update_firmware.md). A floating input can still pick up environmental noise, so verify acquisition with a known high or low level or an actual signal instead of using a floating channel as a reference.
|
||||
|
||||
**Q:**When sampling waveforms, the waveforms in the display window do not match the actual waveforms.
|
||||
|
||||
@@ -58,7 +63,7 @@ The following lists some common problems according to the form of Q&A, please tr
|
||||
|
||||
**Q:**The serial port function in DAPLink mode does not work properly and messages are echoed back
|
||||
|
||||
**A:**Please refer [here](./update_firmware.md) to update the firmware to the latest version before trying again
|
||||
**A:**This issue is fixed in `slogic_combo8_pack_202608121500.bin`. [Update to this release or a newer one](./update_firmware.md). The new release improves sustained high-speed transfers, serial-setting changes, and recovery after USB reconnection.
|
||||
|
||||
**Q:**DAPLink error while burning firmware:`Connection refused due to device mismatch!`
|
||||
|
||||
@@ -76,4 +81,3 @@ None
|
||||
**Q:**The serial terminal cannot open the serial port.
|
||||
|
||||
**A:**Try using a different serial terminal. For Windows, you can try using XCOM, and for Linux, you can try using Minicom.
|
||||
|
||||
|
||||
@@ -2,6 +2,11 @@
|
||||
title: Update Firmware
|
||||
keywords: LogicAnalyzer, debugger, link, tool
|
||||
update:
|
||||
- date: 2026-08-12
|
||||
version: v0.2
|
||||
author: Sipeed
|
||||
content:
|
||||
- Added the fixes in `slogic_combo8_pack_202608121500.bin` and post-upgrade checks
|
||||
- date: 2023-09-01
|
||||
version: v0.1
|
||||
author: lxo
|
||||
@@ -17,10 +22,15 @@ Tool: [Click to download](https://dl.sipeed.com/shareURL/SLogic/SLogic_combo_8/4
|
||||
|
||||
Firmware: [Click to download](https://dl.sipeed.com/shareURL/SLogic/SLogic_combo_8/4_application/Firmware)
|
||||
|
||||
Just select the latest version of the burning tool and firmware, and unzip it after downloading.
|
||||
Use the latest flashing tool. Select `slogic_combo8_pack_202608121500.bin` or a newer firmware release, then extract the downloaded files.
|
||||
|
||||
> Note:When the firmware is named `slogic_combo8_pack_202308171404.bin`, the date is 17/08/2023. The date naming rules are similar for other firmware.
|
||||
|
||||
`slogic_combo8_pack_202608121500.bin` includes these fixes:
|
||||
|
||||
- LA: fixes the dirty waveform on a floating CH7.
|
||||
- DAPLink: fixes serial instability and improves sustained high-speed transfers, serial-setting changes, and recovery after USB reconnection.
|
||||
|
||||
## Configure Tool
|
||||
|
||||
1. Start the tool
|
||||
@@ -59,4 +69,8 @@ Configure the serial port and baud rate, and click `Create & Download` to downlo
|
||||
|
||||

|
||||
|
||||
After the download is completed, the progress bar displays a green box, indicating that the download is successful and the firmware update is completed.
|
||||
When the progress bar turns green, power-cycle the device and check the update:
|
||||
|
||||
1. Switch to blue logic-analyzer mode and confirm that the system detects `SLogic8 U2`.
|
||||
2. Switch to green DAPLink mode and confirm that the system detects `RV CMSIS-DAP`.
|
||||
3. If the update was intended to resolve DAPLink serial or debug stability, repeat the original test with the same wiring and baud rate.
|
||||
|
||||
@@ -2,6 +2,11 @@
|
||||
title: Using as DAPLink
|
||||
keywords: LogicAnalyzer, debugger, link, RISCV, tool
|
||||
update:
|
||||
- date: 2026-08-12
|
||||
version: v0.2
|
||||
author: Sipeed
|
||||
content:
|
||||
- Documented the DAPLink serial stability fix
|
||||
- date: 2023-07-26
|
||||
version: v0.1
|
||||
author: ctx
|
||||
@@ -32,10 +37,12 @@ Here's a step-by-step guide on using the STM32F103C8T6 chip with Windows MDK IDE
|
||||
|
||||
In DAPLink mode, it can simultaneously support one DAPLink interface and one UART interface.
|
||||
|
||||
The pins on the left side of the diagram (TXD, RXD) can be used as a UART interface. (note: DTR and RTS are reserved.)
|
||||
The pins on the left side of the diagram (TXD, RXD) carry UART data. The latest firmware also outputs the CDC DTR and RTS control signals.
|
||||
|
||||
The pins on the right side of the diagram (TCK, TDI, TDO, TMS) are used for DAPLink debugging.
|
||||
|
||||
`slogic_combo8_pack_202608121500.bin` fixes DAPLink serial instability and improves sustained high-speed transfers, serial-setting changes, and recovery after USB reconnection. If data is lost, unexpected echoes appear, or the serial bridge stops transferring, [update to this release or a newer one](./update_firmware.md).
|
||||
|
||||
### Connecting DAPLink using MDK
|
||||
- Set the Reset Options to SYSRESETREQ.
|
||||
- In CMSIS-DAP, locate and select the RV CMSIS-DAP device.
|
||||
@@ -51,4 +58,3 @@ The pins on the right side of the diagram (TCK, TDI, TDO, TMS) are used for DAPL
|
||||

|
||||
|
||||
Thank you for the document contribution from 'dragonforward'.
|
||||
|
||||
|
||||
@@ -2,6 +2,11 @@
|
||||
title: Using as a Logic Analyzer
|
||||
keywords: LogicAnalyzer, debugger, link, RISCV, tool
|
||||
update:
|
||||
- date: 2026-08-12
|
||||
version: v0.2
|
||||
author: Sipeed
|
||||
content:
|
||||
- Updated the device name and removed the resolved floating-CH7 note
|
||||
- date: 2023-07-26
|
||||
version: v0.1
|
||||
author: ctx
|
||||
@@ -16,7 +21,7 @@ Press the button to switch the indicator light to blue. As shown below:
|
||||

|
||||
|
||||
> To verify if Slogic functionality is enabled:
|
||||
> Open the device manager in Windows environment, and use the lsusb command in Linux environment to find the "USB TO LA" device
|
||||
> Check Device Manager on Windows or run `lsusb` on Linux. The device name is `SLogic8 U2`.
|
||||
|
||||
## Getting Started
|
||||
|
||||
@@ -93,8 +98,6 @@ The connection will be made automatically on first startup, or you can choose to
|
||||
|
||||

|
||||
|
||||
> Note:If the D7 channel is not used during the sampling process, a level inversion of the channel may be observed, but this is normal and does not affect normal use.
|
||||
|
||||
#### Analyze the Sampling Result
|
||||
|
||||
1. Open the protocol analysis function
|
||||
|
||||
@@ -2,6 +2,11 @@
|
||||
title: 常见问题(FAQ)
|
||||
keywords: LogicAnalyzer, debugger, link, RISCV, tool
|
||||
update:
|
||||
- date: 2026-08-12
|
||||
version: v0.2
|
||||
author: Sipeed
|
||||
content:
|
||||
- 更新 CH7 悬空波形和 DAPLink 串口问题
|
||||
- date: 2023-09-12
|
||||
version: v0.1
|
||||
author: lxo
|
||||
@@ -31,9 +36,9 @@ update:
|
||||
|
||||
**A:**这可能是接触不稳定导致设备断连了,尝试重新插拔设备后再重新连接即可
|
||||
|
||||
### 使用8通道采集时发现D7通道在悬空时也会出现波形
|
||||
### 使用 8 通道采集时发现 CH7 悬空波形不干净
|
||||
|
||||
**A:**这是一个待解决的问题,但这个问题不会影响采集波形,给D7通道接入外部信号后仍然可以正常采集波形。如果不需要这个通道时也可以隐藏这个通道,很抱歉给您带来了困扰。
|
||||
**A:**该问题已在 `slogic_combo8_pack_202608121500.bin` 中修复,请[更新到该版本或更新版本](./update_firmware.md)。悬空输入仍可能拾取环境噪声;判断采集是否正常时,应连接确定的高、低电平或实际待测信号,不要以悬空通道为基准。
|
||||
|
||||
### 波形采集时,波形显示窗口中的波形和实际的波形不一致
|
||||
|
||||
@@ -57,7 +62,7 @@ update:
|
||||
|
||||
### DAPLink模式的串口功能无法正常使用,出现回显的情况
|
||||
|
||||
**A:**请在更新固件章节中将固件更新到最新版本后再尝试
|
||||
**A:**该问题已在 `slogic_combo8_pack_202608121500.bin` 中修复,请[更新到该版本或更新版本](./update_firmware.md)。新版改进了高速连续传输、串口参数切换以及 USB 重新连接后的恢复过程。
|
||||
|
||||
### DAPLink在烧录固件时报错`Connection refused due to device mismatch!`
|
||||
|
||||
@@ -81,4 +86,3 @@ None
|
||||
### 串口终端无法打开串口
|
||||
|
||||
**A:**尝试一下其他的串口助手,Windows上可以尝试使用XCOM,Linux上可以尝试使用Minicom
|
||||
|
||||
|
||||
@@ -2,6 +2,11 @@
|
||||
title: 更新固件
|
||||
keywords: LogicAnalyzer, debugger, link, tool
|
||||
update:
|
||||
- date: 2026-08-12
|
||||
version: v0.2
|
||||
author: Sipeed
|
||||
content:
|
||||
- 增加 `slogic_combo8_pack_202608121500.bin` 修复说明和升级后检查项
|
||||
- date: 2023-09-01
|
||||
version: v0.1
|
||||
author: lxo
|
||||
@@ -17,10 +22,15 @@ update:
|
||||
|
||||
固件:[点击下载](https://dl.sipeed.com/shareURL/SLogic/SLogic_combo_8/4_application/Firmware)
|
||||
|
||||
烧录工具和固件选择最新版本即可,下载完成后自行解压
|
||||
烧录工具选择最新版本。固件请选择 `slogic_combo8_pack_202608121500.bin` 或更新版本,下载完成后自行解压。
|
||||
|
||||
> 注:当固件名为`slogic_combo8_pack_202308171404.bin`时,日期为2023年08月17日。其他固件的日期命名规则类似。
|
||||
|
||||
`slogic_combo8_pack_202608121500.bin` 包含以下修复:
|
||||
|
||||
- LA:修复 CH7 悬空时波形不干净的问题;
|
||||
- DAPLink:修复串口不稳定问题,改进高速连续传输、串口参数切换以及 USB 重新连接后的恢复过程。
|
||||
|
||||
## 配置烧录工具
|
||||
|
||||
1. 启动烧录工具
|
||||
@@ -59,4 +69,8 @@ update:
|
||||
|
||||

|
||||
|
||||
下载完成后,进度条显示绿框说明下载成功,更新固件完成
|
||||
下载完成后,进度条显示绿框说明下载成功。重新上电,并按以下步骤确认更新结果:
|
||||
|
||||
1. 切换到蓝灯逻辑分析仪模式,确认系统识别到 `SLogic8 U2`;
|
||||
2. 切换到绿灯 DAPLink 模式,确认系统识别到 `RV CMSIS-DAP`;
|
||||
3. 如果升级是为了解决 DAPLink 串口或调试稳定性问题,请按原连接和速率重新测试。
|
||||
|
||||
@@ -2,6 +2,11 @@
|
||||
title: 作为DAPLink使用
|
||||
keywords: LogicAnalyzer, debugger, link, RISCV, tool
|
||||
update:
|
||||
- date: 2026-08-12
|
||||
version: v0.2
|
||||
author: Sipeed
|
||||
content:
|
||||
- 补充 DAPLink 串口稳定性修复
|
||||
- date: 2023-07-23
|
||||
version: v0.1
|
||||
author: lxo
|
||||
@@ -32,10 +37,12 @@ SLogic Combo 8 具备高速DAP-Link功能,适用于各种常规芯片,尤其
|
||||
|
||||
DAPLink模式下可以同时支持一路DAPLink和一路UART
|
||||
|
||||
- 上图左侧的引脚(TXD、RXD)可以作为串口使用(注:DTR、RTS 保留未使用)
|
||||
- 上图左侧的引脚(TXD、RXD)用于串口收发。最新固件也会同步输出 CDC DTR、RTS 控制信号。
|
||||
|
||||
- 上图右侧的引脚(TCK(SWCLK)、TDI、TDO、TMS(SWDIO))作为DAPLink调试使用
|
||||
|
||||
`slogic_combo8_pack_202608121500.bin` 修复了 DAPLink 模式下的串口不稳定问题,改进了高速连续传输、串口参数切换以及 USB 重新连接后的恢复过程。遇到丢数据、异常回显或串口停止收发时,请先[更新到该版本或更新版本](./update_firmware.md)。
|
||||
|
||||
### 使用MDK连接DAPLink
|
||||
|
||||
1. 设置Reset Options为SYSRESETREQ
|
||||
@@ -82,4 +89,4 @@ xpack 版本 openocd 下载链接为https://xpack.github.io/dev-tools/openocd/re
|
||||

|
||||
|
||||
|
||||
感谢群友`dragonforward`的文档贡献
|
||||
感谢群友`dragonforward`的文档贡献
|
||||
|
||||
@@ -2,6 +2,11 @@
|
||||
title: 作为逻辑分析仪使用
|
||||
keywords: LogicAnalyzer, debugger, link, RISCV, tool
|
||||
update:
|
||||
- date: 2026-08-12
|
||||
version: v0.2
|
||||
author: Sipeed
|
||||
content:
|
||||
- 更新设备名称并移除已修复的 CH7 悬空通道异常说明
|
||||
- date: 2023-07-23
|
||||
version: v0.1
|
||||
author: lxo
|
||||
@@ -16,7 +21,7 @@ update:
|
||||

|
||||
|
||||
> 验证逻辑分析仪功能是否开启:
|
||||
> Windows环境打开设备管理器,Linux环境使用lsusb命令,可以找到"USB TO LA"设备
|
||||
> Windows 环境可在设备管理器中查看,Linux 环境可运行 `lsusb`,设备名称为 `SLogic8 U2`。
|
||||
|
||||
## 开始使用
|
||||
|
||||
@@ -91,8 +96,6 @@ PulseView-SLogic-250910-aarch64.dmg
|
||||
|
||||

|
||||
|
||||
> 若在采样过程中未使用D7通道,可能会观察到该通道的电平反转现象,该现象为正常情况不影响正常使用
|
||||
|
||||
#### 分析采样结果
|
||||
|
||||
1. 打开协议选择栏
|
||||
@@ -230,4 +233,4 @@ PulseView-SLogic-250910-aarch64.dmg
|
||||
1. 逻辑分析仪与电脑是共地的,如果待测设备是强电系统,务必使用"USB隔离器"来进行隔离措施。否则,很可能会造成逻辑分析仪或电脑损坏的风险
|
||||
2. GND 通道与被测系统的 GND 必须可靠连接,尽可能的短
|
||||
3. 信号通道必须可靠连接至被测系统的待测信号位置,不可随意“嫁接”,导致干扰的引入
|
||||
4. 如果不注意接线方式,很可能会引入很多毛刺,导致软件无法分析数据
|
||||
4. 如果不注意接线方式,很可能会引入很多毛刺,导致软件无法分析数据
|
||||
|
||||